Automated Genomic DNA Extraction

互联网

712

实验试剂

 

ChargeSwitch® Lysis Buffer (L12)

Proteinase K

ChargeSwitch® Magnetic Beads

ChargeSwitch® Purification Buffer (N5)

ChargeSwitch® Wash Buffer (W12)

ChargeSwitch® Elution Buffer (E5) or TE Buffer (not supplied; 10 mM Tris-HCl, 1 mM EDTA, pH 8.5)

实验设备

 

Liquid handling robot configured to process samples in 96-well plates

96 x 2 ml deep well plate(s)

96 x 300 µl U-bottomed microtiter plate

实验材料

 

10-20 µl blood samples

实验步骤

 

Before Starting
Perform the following before beginning:

Automated Protocol
Follow the protocol below to isolate genomic DNA from 10-20 µl blood samples. The volumes given are on a per sample basis.

  1. Start with 96 x 10-20 µl blood samples in a 96 x 2 ml deep well plate.
  2. Add 500 µl of Lysis Mix and incubate at room temperature for 10 minutes. Once during the incubation, pipet up and down gently 15 times to mix. Set the pipette tip to 350 µl and avoid forming bubbles.
  3. Add 70 µl of Purification Mix (make sure that the beads are thoroughly resuspended)
  4. Shake at medium fast speed (e.g. pulse, 10 seconds) to evenly distribute the magnetic beads within the solution.
  5. Shake samples rapidly for 20 seconds to mix.
  6. Wait for 30 seconds.
  7. Move samples to the 96-Well Magnetic Separator.
  8. Wait for 90 seconds.
  9. Slowly aspirate all of the supernatant and discard, leaving behind the pellet of beads.
  10. Remove samples from the 96-Well Magnetic Separator.
  11. Add 500 µl of ChargeSwitch® Lysis Buffer (L12; no Proteinase K) and shake samples rapidly for 20 seconds to evenly distribute the magnetic beads within the solution.
  12. Add 50 µl of ChargeSwitch® Purification Buffer (N5) and shake at medium speed for 20 seconds to mix. Samples should appear clear, with no brown flecks.
  13. Wait for 30 seconds.
  14. Move samples to the 96-Well Magnetic Separator.
  15. Wait for 60 seconds.
  16. Slowly aspirate all of the supernatant and discard, leaving behind the pellet of beads.
  17. Remove samples from the 96-Well Magnetic Separator.
  18. Add 500 µl of ChargeSwitch® Wash Buffer (W12).
  19. Shake at medium speed (e.g. pulse, 10 seconds) to evenly distribute the magnetic beads within the solution.
  20. Move samples to the 96-Well Magnetic Separator.
  21. Wait for 60 seconds.
  22. Slowly aspirate all of the supernatant and discard, leaving behind the pellet of beads.
  23. Leave samples on the 96-Well Magnetic Separator for the second wash.
  24. Add 500 µl of ChargeSwitch® Wash Buffer (W12).
  25. Wait for 30-60 seconds.
  26. Slowly aspirate all of the supernatant and discard, leaving behind the pellet of beads.
  27. Move samples to the shaker.
  28. Add 100 µl of Elution Buffer. Pipet up and down gently 50 times to mix (set the pipette tip to 75 µl).
  29. Shake rapidly for 1-2 minutes to completely disperse the beads within the solution.
  30. Move samples to the 96-Well Magnetic Separator.
  31. Wait for 1 minute.
  32. Slowly aspirate supernatant containing the DNA to a 96 x 300 µl U-bottomed microtiter plate.

Storing DNA
Store the purified DNA at -20°C or use immediately for downstream analysis. Avoid repeatedly freezing and thawing DNA.
Quantitating DNA Yield
To quantitate yield of your DNA, use the Quant-iT™ PicoGreen® dsDNA Quantitation Kit (Catalog no. P7589).

注意事项

 

To maximize DNA yield, follow these recommendations when processing your samples:

 

 