丁香实验_LOGO
登录
提问
我要登录
|免费注册
点赞
收藏
wx-share
分享

Amicon Stirred Ultrafiltration Cells

互联网

2140

Amicon Stirred Ultrafiltration Cells
(Models 8050, 8400)
Description - Membranes - Assembly - Operation - Limits - Sterilization - Maintenance - Troubleshooting


Description

For protein concentration, gas pressure is applied directly to ultrafiltration cell. Solutes above the membrane's molecular weight (MW) cut-off are retained in cell, while water and solutes below the cut-off pass into the filtrate and out of cell.
 
  8050 8400
Cell capacity (ml)
Stirred minimum volume¹ (ml)
Membrane diameter (mm)
Effective membrane area (cm ²)
Hold-up volume² (ml)
50
2.5
43
13.4
0.5
400
10.0
76
41.8
1.5
Maximum operating pressure: 75 psi (5.3 kg/cm²)
Relief valve setting: 90 psi (6.3 kg/cm²)
Maximum diafiltration operating pressure: 55 psi (3.9 kg/cm²)
¹ Minimum working volume.
² Non-recoverable volume (below membrane surface).

Membranes


Assembly

  1. Place membrane in holder, shiny side up ; then place O-ring on top of membrane. Gently push O-ring down so that it contacts and seats membrane evenly in bottom of holder. Handle membrane by its edges to avoid scratching or contaminating surface.
  2. Fit membrane holder into cell body, aligning tabs on sides of holder with slots in base of cell body.
  3. Invert cell body and membrane holder; screw base firmly into bottom of cell body. A definite "stop" will be felt when base and body are fully engaged. Top of membrane holder will be flush with bottom of slots in cell body.
  4. Push filtrate exit tubing onto exit spout of membrane support.
  5. Place stirrer assembly into cell body. When properly installed, arms of stirrer assembly will rest on small ridge inside top of cell body.
  6. Introduce sample into cell.
  7. With a twisting motion, push cap down onto cell body, orienting gas inlet port on cap opposite filtrate exit port on holder. If cap assembly does not slide easily, lubricate O-ring lightly with petroleum jelly. Do not allow petroleum jelly to contact membrane!
  8. Set pressure-relief valve to horizontal (open) position.
  9. Slide cell into retaining stand, fitting ring on cell base into hole in stand. Flattened edges on bottom flange of cap ensures that cell is inserted properly and prevents rotation of cell once inside stand.
  10. Turn pressure-relief valve to vertical (closed) position.
  11. Attach gas pressure line.

Operation

  1. Place cell on magnetic stirring table.
  2. Connect inlet line to regulated gas pressure source (Nitrogen gas is recommended for pressurizing cell! Use of compressed air can cause large pH shifts due to dissolution of carbon dioxide. With sensitive solutions, oxidation can also occur, leading to other potential problems.)
  3. Hold cell steady on the stirring table and pressurize according to instructions in membrane package. Generally 55 psi (3.7 kg/cm2) is optimal, maximal 70 psi (4.7 kg/cm2) nitrogen gas pressure. Cap assembly moves upward, forming a secure lock with retaining stand once system is pressurized.
  4. Turn on stirring table and adjust stirring rate until the vortex created is approximately one-third the depth of the liquid volume.
  5. When finished, turn off nitrogen pressure source and stirring table.
  6. Vent pressure inside cell by slowly turning pressure-relief knob to horizontal position. Push cap down, then slide cell out from retaining stand. Note : Overly rapid depressurization can cause the membrane to buckle up and rupture.
  7. Using a twisting motion, remove cell cap and the magnetic stirrer assembly. Always remove the cell top with the pressure-relief valve set to the horizontal (open) position. Removal in vertical (closed) position can create a partial vacuum which can rupture the membrane! Pour out solution.
  8. Disassemble cell, then wash all components with a mild detergent/water solution, then rinse thoroughly. Caustic cleaning solutions may damage the anodized aluminum retaining stand. Leave ultrafiltration cell disassembled whenever it is unlikely to be used for several weeks.

Limits


Sterilization

Amicon stirred ultrafiltration cells can be autoclaved at 121°C (250°F) for 30 minutes. They are compatible with standard sterilizing gas mixtures, 70% ethanol, isopropanol or 5% formalin. CAUTION: Tighten the base partially before autoclaving.

Maintenance


Troubleshooting

Little or no filtrate obtained: Filtrate rate abnormally high: Cell leaks:

 

提问
扫一扫
丁香实验小程序二维码
实验小助手
丁香实验公众号二维码
扫码领资料
反馈
TOP
打开小程序