MRTHOD:
-
Place sample (containing 0.5 - 10 µg GlcN) in a Pyrex screw capped tube.
-
Add HCl to a final concentration of 2N and a final volume of 0.6 ml. Flush with nitrogen.
-
Stopper tightly and hydrolyze 16h at 100o C.
-
Prepare standards containing 0 - 20 µg GlcN from a stock solution of 1.5 mg/ml. (Use 0, 3, 5, 8, 10, 12 ml of stock solution and add water to make up to 300 µl. Then add 300 µl 4N HCl to a final volume of 0.6 ml and final concentration of 2N). Standards need not be heated.
-
Neutralize samples and standards with 0.4 ml 2M Na2 CO3 (10.6 g in 50 ml). (pH - 7 with the slight excess of Na2 CO3 )
-
Shake gently and add 0.5 ml (freshly prepared) 2% acetyl acetone in 1.5M Na2 CO3 . (15.9 g Sodium Carbonate + 2 ml acetyl acetone made up to 100 ml).
-
Stopper tightly and heat in boiling water bath for 20 min.
-
Cool. Add 1 ml EtOH.
-
Add 0.5 ml Ehrlichs reagent. (1 g p dimethylaminobenzaldehyde in 15 ml EtOH and 15 ml c/HCl)
-
Shake tubes vigorously to expel excess CO2 .
-
Maximum colour development is reached in 5 min. Chromophore stable for 1 to 2 hours.
- Read absorbance at 530 nm.
NOTE: NaCl affects colour. Therefore standards and sample should contain the same amount of salt to avoid colour depression and erroneous results. Dialysis helps to remove NaCl.